Leeb portable hardness tester MET-D1
Technical characteristics
| HARDNESS SCALE | MEASUREMENT RANGE |
LIMITS OF ALLOWABLE ABSOLUTE HARDNESS TESTER ERROR |
|---|---|---|
| Rockwell "C" scale | (20 - 70) HRC | ± 2 HRC |
| Brinell scale | (75 - 450) HB | ± 12 HB |
| Vickers scale | (75 - 1000) HV | ± 15 HV |
| Shore scale | (23 - 102) HSD | ± 3HSD |
| MEASUREMENT PRINCIPLE | Rebound (Leib) principle | |
|---|---|---|
| PROBES TYPES | D, DK | |
| DIAMETER OF CARBIDE BALL | not less than 2.5 mm | |
| ROUGHNESS OF SAMPLE SURFACE | not more than 3.2 Ra | |
| RADIUS OF CURVATURE OF SAMPLE SURFACE | not less than 10 mm | |
| MINIMUM WEIGHT OF SAMPLE | not less than 3 kg | |
| SAMPLE THICKNESS | not less than 12 mm | |
| PROBE SERVICE LIFE | Not less than 100,000 measurements | |
| POWER SUPPLY | AC mains, V/Hz | 100-240/50-60 |
| Battery | 1.2 V | |
| Power consumption | >3.0 VA | |
| BATTERY CHARGING TIME | 8 hours | |
| BATTERY WORKING TIME | Without backlight | 16 h |
| With backlight | 5 h | |
| From AC power | unlimited | |
| OPERATING TEMPERATURE | from -10° to +50°С | |
| TRANSPORTATION AND STORAGE TEMPERATURE | from -35° to +60°С | |
| OPERATING CONDITIONS | Relative humidity 30 to 80% | |
| OVERALL SIZE | ||
| MET-D1 electronic unit | 145х80х40 mm | |
| Probe D (length x diameter) | 140x25 mm | |
| Probe DSh (length x diameter) | 80x25 mm | |
| HARDNESS TESTER MASS (UNIT WITH PROBE) | ||
| Hardness tester MET-D1 | 0.45 kg | |
| PROCESSING MEASUREMENT RESULTS | Calculating the average value, deleting the current measurements, data saving, working with archive. | |
| DISPLAY BACKLIGHT | Available | |
| DISPLAY FEATURES | Hardness scale, measured value, number of measurements, operating mode, archive number, battery charge indicator, auto power off after 150 sec. | |
| MEMORY (ARCHIVE) | 99 readings that are saved when the power is turned off | |
| INTERNAL SOFTWARE "MET-D" (OR "MET-DA") | Available. Allows you to work with the hardness tester without a connection to a computer. | |
| EXTERNAL SOFTWARE "MET" | Available. Transfers data from the hardness tester archive to the computer. Allows you to process measurement data, create databases, build graphs, print results. | |
| TIME OF ONE MEASUREMENT, S | 1.5 | |
Advantages
Package
| NAME | |
|---|---|
| Quantity | |
| Electronic unit MET-D1 in a plastic case | 1 |
| Dynamic probe D or short probe DSh (optional) | 1 |
| Pusher for probe DSh (when selecting probe DSh) | 1 |
| Cable for connecting the dynamic probe to the electronic unit | 1 |
| USB cable for connection to computer and charger | 1 |
| Charger | 1 |
| Battery size C | 1 |
| Key for battery compartment screws | 1 |
| CD with external software "MET" and operating manual MET.39601863.009 RE | 1 |
| Verification procedure MET-TP.01 MP | 1 |
| Verification certificate | 1 |
| Passport | 1 |
| Universal case for fixing the electronic unit on the forearm, neck and belt | 1 |
| Case | 1 |
Calibration, verification, certification
The initial verification of hardness testers is carried out when the hardness testers are released from production. A "Verification Certificate" of the established form is issued for the hardness testers, which is included in the hardness tester delivery set.
Verification and calibration of portable hardness testers of the "MET" series is carried out according to the main hardness scales HRC, HB, HV, HSD.
Calibration according to additional (user) scales H1, H2, H3 (for example, according to the hardness scales HBW, HRB, HRN, HRT, etc.), as well as for metals that differ significantly in properties from steel (for example, aluminum and copper alloys, etc.) is carried out by the user according to the instructions provided in the hardness tester passport. The Rm scale of tensile strength is available, which is calculated from the HB scale according to GOST 22761 (Russian Federation).
The frequency of verification of hardness testers is once a year. Verification can be carried out by any metrological service of legal entities accredited for the right to verify hardness measuring instruments, regional Centers for Standardization and Metrology (CSM). The verification method together with the Passport are included in the hardness tester kit.